Fluke 773 Milliamp Process Clamp Meter
Specifications: Fluke 773 Milliamp Process Clamp Meter
Specifications |
Diameter of measurable conductor |
0.177" or 4.5 mm max |
Operating temperature |
-10 to 50ºC |
Storage temperature |
-25 to 60ºC |
Operating humidity |
< 90% @ <30ºC, < 75% @ 30 to 55ºC |
Operating altitude |
0 to 2000 m |
IP rating |
IP 40 |
Size |
44 x 70 x 246 mm (2 7/8 x 5 ¾ x 11 5/8 in) |
Weight |
410 g, (14.4 oz) |
Vibration |
Random 2 g, 5 to 500 Hz |
Shock |
1 meter drop test (except the jaw) |
EMI/RFI |
Meets EN61326-1 |
For current measurement w/ JAW, add 1mA to specification for EMC field strengths of 1V/m up to 3V/m. |
Temperature coefficient |
0.01% ºC
(< 18ºC or > 28ºC) |
Power, battery life |
(4) AA 1.5 V Alkaline, IEC LR6, 12 hours in mA source into 500 ohms |
Warranty |
Three years for electronics
One year for cable and clamp assembly |
Functional Specifications |
mA measurement
(measured by clamp) |
Resolution / range |
0 to 20.99 mA |
Accuracy |
0.2% + 5 counts |
Resolution / range |
21.0mA to 100.0 mA |
Accuracy |
1% + 5 counts |
mA measurement (measured in series with test jacks) |
Resolution / range |
0 to 24.00 mA |
Accuracy |
0.2% + 2 counts |
mA source (maximum mA drive: 24 mA into 1,000 ohms) |
Resolution / range |
0 to 24.00 mA |
Accuracy |
0.2% + 2 counts |
mA simulate (maximum voltage 50 V dc) |
Resolution / range |
0 to 24.00 mA |
Accuracy |
0.2% + 2 counts |
Voltage source (2 mA maximum drive current) |
Resolution / range |
0 to 10.00 V DC |
Accuracy |
0.2% + 2 counts |
Voltage measure |
Resolution / range |
0 to 30.00 V DC |
Accuracy |
0.2% + 2 counts |