Product characteristics
| Number of inputs and outputs | Number of digital outputs: 2; Number of analogue outputs: 1 | |||
| Measuring range |
|
|||
| Process connection | threaded connection G 1 external thread Aseptoflex Vario |
Application
| Special feature | Gold-plated contacts | |||
| Application | flush mountable for the food and beverage industry | |||
| Media | viscous media and liquids with suspended particles; liquids and gases | |||
| Medium temperature[°C] | -25…150 | |||
| Min. bursting pressure |
|
|||
| Pressure rating |
|
|||
| Vacuum resistance |
|
|||
| Type of pressure | relative pressure; vacuum | |||
| No dead space | yes | |||
| MAWP for applications according to CRN[bar] | 10 |
Electrical data
| Min. insulation resistance[MΩ] | 100; (500 V DC) |
| Protection class | III |
| Reverse polarity protection | yes |
| Integrated watchdog | yes |
| 2-wire | |
|---|---|
| Operating voltage[V] | 20…30 DC |
| Current consumption[mA] | 3.5…21.5 |
| Power-on delay time[s] | < 1 |
| 3-wire | |
|---|---|
| Operating voltage[V] | 18…30 DC |
| Current consumption[mA] | 5…45; (430 bei max. Laststrom) |
| Power-on delay time[s] | < 0.5 |
Inputs / outputs
| Number of inputs and outputs | Number of digital outputs: 2; Number of analogue outputs: 1 |
Outputs
| Total number of outputs | 2 |
| Output signal | switching signal; analogue signal; IO-Link |
| Electrical design | PNP/NPN |
| Number of digital outputs | 2 |
| Output function | normally open / normally closed; (parameterisable) |
| Number of analogue outputs | 1 |
| Analogue current output[mA] | 4…20, invertible; (scalable) |
| Short-circuit protection | yes |
| Type of short-circuit protection | pulsed |
| Overload protection | yes |
| 2-wire | |
|---|---|
| Max. load[Ω] | 300 |
| 3-wire | |
|---|---|
| Max. voltage drop switching output DC[V] | 2 |
| Permanent current rating of switching output DC[mA] | 100 |
| Switching frequency DC[Hz] | 125 |
| Max. load[Ω] | (Ub – 10 V) / 21,5 mA; 650 Ω (Ub = 24 V) |
Measuring/setting range
| Measuring range |
|
||||||||||||
| Set point SP |
|
||||||||||||
| Reset point rP |
|
||||||||||||
| Analogue start point |
|
||||||||||||
| Analogue end point |
|
||||||||||||
| Min. difference between SP and rP |
|
||||||||||||
| In steps of |
|
||||||||||||
| Factory setting |
|
| Temperature monitoring | |||
|---|---|---|---|
| Measuring range |
|
||
Accuracy / deviations
| Switch point accuracy[% of the span] | < ± 0,2; (DIN EN IEC 62828-1; Turn down 1:1) | ||||||||
| Repeatability[% of the span] | < ± 0,1; (with temperature fluctuations < 10 K; Turn down 1:1) | ||||||||
| Characteristics deviation[% of the span] | < ± 0,2; (DIN IEC EN 62828-1 incl. zero point and span error, non-linearity, hysteresis; Turn down 1:1) | ||||||||
| Linearity deviation[% of the span] | < ± 0,15; (Turn down 1:1) | ||||||||
| Zero-point stabilisation[% of the span] |
|
||||||||
| Hysteresis deviation[% of the span] | < ± 0,15; (Turn down 1:1) | ||||||||
| Long-term stability[% of the span] | < ± 0,1; (Turn down 1:1; per year) | ||||||||
| Total deviation over temperature range |
|
||||||||
| Notes on the accuracy / deviation | for further details see section Diagrams and graphs |
| Temperature monitoring | |
|---|---|
| Accuracy[K] | ± 2,5+ (0,08 x ( Umgebungstemperatur – Mediumtemperatur )) |
| Repeatability[K] | ± 0,2 |
| Resolution[K] | 0.2 |
Response times
| Damping process value dAP[s] | 0…99.99 |
| Damping for the analogue output dAA[s] | 0…99.99 |
| 2-wire | |
|---|---|
| Step response time analogue output[ms] | 30 |
| 3-wire | |
|---|---|
| Min. response time of switching output (dAP)[ms] | 3 |
| Step response time analogue output[ms] | 7 |
Interfaces
| Communication interface | IO-Link | ||||||||||
| Transmission type | COM2 (38,4 kBaud) | ||||||||||
| IO-Link revision | 1.1 | ||||||||||
| SDCI standard | IEC 61131-9 | ||||||||||
| Profiles |
|
||||||||||
| SIO mode | yes | ||||||||||
| Required master port type | A | ||||||||||
| Min. process cycle time[ms] | 5.6 | ||||||||||
| IO-Link resolution pressure[bar] | 0.0005 | ||||||||||
| IO-Link resolution temperature[K] | 0.2 | ||||||||||
| IO-Link process data (cyclical) |
|
||||||||||
| IO-Link functions (acyclical) | application specific tag; internal temperature; operating hours counter; switching cycles counter; Pressure peak counter; measuring cell diagnosis | ||||||||||
| Supported DeviceIDs |
|
Operating conditions
| Ambient temperature[°C] | -25…80 |
| Storage temperature[°C] | -40…100 |
| Protection | IP 67; IP 68; IP 69K |
Tests / approvals
| EMC |
|
||||
| Shock resistance |
|
||||
| Vibration resistance |
|
||||
| MTTF[years] | 214 | ||||
| UL approval |
|
Mechanical data
| Weight[g] | 360 |
| Housing | cylindrical |
| Dimensions[mm] | Ø 50 / L = 107 |
| Materials | Housing: stainless steel (316L/1.4404); pushbuttons: PBT; sealings: FKM; M12 connector: PEI; Display: PFA; ventilation diaphragm: PTFE |
| Materials (wetted parts) | measuring cell: ceramics (99.9 % Al2O3); Process connection: stainless steel (316L/1.4435) surface characteristics: Ra < 0,4 µm / Rz = 4 µm; Sealing: PTFE |
| Min. pressure cycles | 100 million |
| Tightening torque[Nm] | 35 |
| Process connection | threaded connection G 1 external thread Aseptoflex Vario |
Displays / operating elements
| Display |
|
||||||||
| Display unit | bar; psi; MPa |
Remarks
| Pack quantity | 1 pcs. |
Electrical connection
| Connection | Connector: 1 x M12; coding: A; Contacts: gold-plated |